96-Well Gradient Quantitative PCR Machine (Cat. No. : FRTTC-96-G)
|
Introduction
With the FRTTC-96-G, we bring gradient real-time quantitative PCR instrumentation to your laboratory with a high-performance, full-featured, and low-cost system. The instrument combines precise thermal control, broad wavelength detection, and powerful software that is easy to learn and use. Most importantly, the FRTTC-96-G system delivers outstanding results with all QRCR chemistries.
|
 |
|
| Innovation Design |
| Nova optics system |
| Unique optical fiber technology |
Exciting the fluorophore and collecting the emission signal from each well independently and simultaneously. |
|
| Signal Collection Technology |
Collecting signal from the side of the tube bottom that prevents the reflection of the excitation light and broadens the type of compatible tube. |
|
| Four-color filter |
Detecting a various of fluorophores in single tube, including FAM, SYBR Green I, TET, HEX,VIC, TAMRA, ROX, Texas Red, Cy3 and Cy5. |
|
|
|
| Thermal cycling |
| 6 independent temperature control modules. |
| Gradient temperature range over 20°C in 12 columns. |
|
|
| Powerful software |
| Powerful data analysis |
Absolute Quantification, Relative Quantification, Multiple Quantification, Mutation Analysis, SNP Analysis, Allelic Analysis, △△Ct Analysis, Dissociation Curve, Standard Sample Curve and Melting Curve Detection will be provided real time, and all data can be re-analyzed. PCR efficiency calculation and calibration well by well, SYBR Green I calibration. |
|
| Powerful data management |
Wide range of personal preferences may be programmed by different operators. |
|
| User friendly interface |
Easily sets up plate parameters and edits the format of your report, and all reports output could be in Excel format. |
|
|
|
|
Features
| 1. |
High-Resolution Melting(HRM). A technique, based on high resolution (0.1oC increment) melting of amplified products in the presence specific fluorescent dsDNA-binding dye, used for mutation scanning and genotyping without labeled probes. |
| 2. |
Short running time. For 5-30ul, 40 cycles can be finished within 40 mins. |
| 3. |
Short readout time. 96 well plate can be detected within 1 seconds. |
| 4. |
Long lifetime of halogen lamp. |
| 5. |
Broad dynamic range. Up to 10 orders of magnitude in starting copy number. |
| 6. |
High sensitivity. To detect single copy of the initial template. |
| 7. |
Peltier-based thermal cycle system enhances the precision of temperature. |
| 8. |
Powerful, versatile software including plate set-up user interface and automated analysis tools make data processing simple and straightforward. |
|
Specifications
|
| Thermal Cycle System |
| Thermal Mode |
Peltier-based TE module |
| Thermal Gradient |
12 Columns |
| Multiple Temperature Control |
6 areas |
| Heating/Cooling Rate |
>3 ℃ /S |
Precision of Temperature Control Homogeneity of Temperature |
±0.1 ℃ ± 0.25 ℃ |
|
| Optical system |
| Excitation Light Source |
Halogen Lamp (>2000hours) |
| Wavelength of Excitation Light |
400-700 nm |
| Wavelength of Detection |
450-750nm |
| Detection Gateway |
4 filter provided |
| Detection System |
Cooled CCD |
| Mode of Detection |
Real Time Dynamic Detecting and Displaying. Four Colors can be detected simultaneously. |
|
| Detection Performance |
| Sensitivity of Detection |
<10 copies of Target Gene Can be Detected at 99.7% Confidence |
| Time of Readout |
96 well plate can be detected within 3 seconds. |
| Time of Sample analysis |
96 well plate can be analyzed within seconds. |
| Order of Magnitude |
10 orders, 100-1010 |
|
| Fluorescent System |
| Fluorescence Report Dyes |
FAMTM/SYBR-Green I, VICTM/JOETM,TAMRATM, ROX, Cy3, Cy5, and Texas Red etc. |
| Fluorescent Quantitative Methods |
Taqman Probe, SYBR Green I Dye, Taqman MGB Probe , and Molecular Beacon etc. |
|
| Sample |
| Sample Format |
0.2ml single tube, 8-strip tube, and 96-well plate |
| Sample Capacity |
96 samples per run |
| Sample Volume |
10-100ul |
| Reaction Vessel |
0.2ml |
| Running Time |
5-30ul, 40cycles can be finished within 40mins |
|
|
|
|
 |
|