AgileReaderTM - All-In-One Fast Microplate Reader
|
|
| Brochure |
Software Manual |
|

Cat. No.: AMPR900 |
Introduction
Using high quality optical components, the AMPR900 Micriplate reader can perform ELISA
measurements between the wavelength of 340~900nm with end point, two point, and
kinetic measurement method. The AMPR900 is built with an incubator and shaker for
temperature sensitive kinetics measurements. To interface with PC, the AMPR900 comes with
the PC-Mate software to control, store, and transfer the analysis results from the
instrument to the PC.
With the versatility and features provided within the AMPR900 Micriplate reader, users can use
this instrument in many different applications such as ELISAs / EIAs, enzymatic activity,
bacterial growth studies, and fast kinetic assays.
|
|
Features
| ¡E |
Multiple measurement modes and analysis methods.
AMPR900 is built with 3 types of measuring method ; end point, two points, and kinetic to
suit the different needs of the users application. To help analyze different results, the
AMPR900 also provides cut-off and quantitative analysis methods.
|
| ¡E |
Flexible and high performance measurement
With the modular filter wheel design of AMPR900, users can customize filter wheel¡¦s
configuration to different wavelength to fit their special applications.
The enhance optical performance of the AMPR900 is also capable of measuring speed lower
than 5s per plate with the upmost precision.
|
| ¡E |
Build-in incubator for kinetic studies.
The AMPR750 Microplate reader is built with digital temperature controller to regulate the
incubator from ambient+3¢J to 50¢J temperature. The built-in advance temperature control makes the AMPR750 ideal for kinetic studies.
|
| ¡E |
PC-Mate software to control and store data through the PC
The convenient PC-Mate software lets users control the measuring method, incubator,
shaker, and analysis method through the PC or notebook in their labs. The PC-Mate
software also transfers the protocols and standard curves of the instrument onto a PC
or notebook for storage.
|
|
Specification
| Cat. No. |
AMPR900 |
| Wavelength range |
340-900 nm |
| Optical system |
9 channels, 1 reference channel |
| Filters |
8 |
| Well type |
96 |
| Measurement range |
0.000~4.000 Abs |
| Measurement time |
<5 sec |
| Measurement modes |
End point, Two Point , Kinetics |
| Accuracy |
+/- 0.005 Abs or 2% |
| Linearity |
+/- 1% from 0.000 to 3.000 Abs |
| Precision |
<0.2% CV from 0.000 to 2.000 Abs
<0.5% CV from 2.000 to 3.000 Abs
|
| Resolution |
0.001 Abs |
| Data storage |
Parameter 100 sets,
Data 50 sets,
Calibration 20 sets |
| Interface |
RS232 or USB |
| Incubator |
+3¢J ~50¢J +/- 0.5¢J |
| Shaking |
8Hz, 11Hz, and 14Hz |
| Dimension |
355 x 345 x 174 (mm) |
| Weight |
14 Kg |
*ACTGene reserves the right to alter specifications without notice.
|
|
Order Information
| AMPR750 Microplate Reader |
AMPR900 |
| Tungsten halogen lamp |
AMPR900-201 |
| Filter cassette 405nm |
AMPR900-301 |
| Filter cassette 450nm |
AMPR900-302 |
| Filter cassette 490nm |
AMPR900-303 |
| Filter cassette 550nm |
AMPR900-304 |
| Filter cassette 570nm |
AMPR900-307 |
| Filter cassette 650nm |
AMPR900-306 |
| Filter cassette 600nm |
AMPR900-305 |
|
|
 |
|
|